4Low Frequency Multiplexer Switch Modules

Low Frequency Multiplexer Switch Modules

All low frequency multiplexer (MUX) switch modules feature two banks of channels that provide broad multiplexing and measuring capabilities. You can connect a MUX to an external instrument, and/or switch multiple analog signals to the internal DMM. With the 34921A, 34922A, 34923A, and the 34924A modules, you can close more than one channel in each bank simultaneously (N:1 configuration). As the 34925A module is protected with overvoltage circuitry, you can close only one channel in each bank at one time (1:N configuration).

And, you can connect multiple MUXes to the built- in Analog Buses, which allow you to scan as many as 560 2- wire (differential) channels or 640 1- wire (single- ended) channels in one 34980A mainframe.

N O TE

Safety Interlock Analog Buses of the 34980A can carry 300 V

signals. MUX and matrix modules with Analog Bus relays have a

 

 

hardware Safety Interlock feature that forces Analog Bus relays

 

open when their associated D-sub connector (faceplate) interlock

 

pins lose continuity. This prevents signals from the Analog Buses

 

from being present on D-sub connector pins. Optional Agilent

 

terminal blocks automatically provide continuity for the interlock

 

pins. When the terminal blocks are not used, you must provide

 

continuity for the interlock pins in the DUT assembly. See pinout

 

drawings and tables in this chapter for the location of interlock pins

 

on the module of interest.

 

MUX modules with Analog Bus connections have Analog Bus

 

relays on each of their two banks. Therefore, the interlock pins are

 

found on both Bank 1 and Bank 2 D-sub connectors of the MUX

 

modules.

 

Normally, if you attempt to connect to the Analog Buses without a

 

terminal block or cable connected, an error is generated. The

 

SYSTem:ABUS:INTerlock:SIMulate command allows

 

you to temporarily disable errors generated by the Safety Interlock

 

feature and enables the Safety Interlock simulation mode. Although

 

Safety Interlock errors are suppressed in this mode, the actual

 

Analog Bus relays affected by the Safety Interlock are disabled as

 

long as no terminal block or cable is connected to the module.

 

 

96

34980A User’s Guide