2Features and Functions

The following table shows the relationship between integration time, measurement resolution, number of digits, and number of bits.

Relationship between integration time, resolution, digits, and bits

Integration Time

Resolution

Digits

Bits

 

 

 

 

0.02 PLC

< 0.0001 x Range

4½ Digits

15

0.2 PLC

< 0.00001 x Range

5½ Digits

18

1 PLC

< 0.000003 x Range

5½ Digits

20

2 PLC

< 0.0000022 x Range

6½ Digits

21

10 PLC

< 0.000001 x Range

6½ Digits

24

20 PLC

< 0.0000008 x Range

6½ Digits

25

100 PLC

< 0.0000003 x Range

6½ Digits

26

200 PLC

< 0.00000022 x Range

6½ Digits

26

 

 

 

 

The instrument selects 1 PLC when the measurement function is changed and after a Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the integration time setting.

Front Panel Operation: DMM or Channel (Configure) > INTEGRATION > TIME

First, select the measurement function on the active channel. You are automatically guided to the next level of the menu where you can select a specific integration time.

Remote Interface Operation: You can set the integration time using the SENSe commands. For example, the following command specifies an aperture time of 2 ms for resistance measurements on channel 2001.

SENS:RES:APER 0.002,(@2001)