Agilent Technologies Switch/Measure manual Front Panel Operation DMM or Channel Measure

Models: Switch/Measure

1 216
Download 216 pages 59.53 Kb
Page 73
Image 73

Features and Functions

2

Front Panel Operation: DMM or Channel (Measure)

For channel monitoring, turn the knob to the desired channel. To stop a Monitor, press the lighted key again.

Remote Interface Operation: Use the following command to select between the channel Monitor mode (default) and the internal DMM monitor mode.

ROUTe:MONitor:MODE {CHANnelDMM}

The following program segment selects the channel to be monitored (specify only one channel) and enables the Monitor function.

ROUTE:MON:CHAN (@1003)

ROUTE:MON:CHAN:ENABLE ON,(@1003)

ROUTE:MON:STATE ON

The following program segment enables the Monitor function on the internal DMM:

ROUTE:MON:MODE DMM

ROUTE:MON:STATE ON

To read the monitor data from the selected channel or the internal DMM, send the following command. Each reading is returned with measurement units, time stamp, channel number, and alarm status information

(see “Reading Format” on page 56).

ROUTe:MONitor:DATA?

34980A User’s Guide

61

Page 73
Image 73
Agilent Technologies Switch/Measure manual Front Panel Operation DMM or Channel Measure