Low Frequency Multiplexer Switch Modules

4

34923A D-Sub Connectors for One-Wire Mode

Bank 1

 

 

 

 

 

 

Bank 1

 

Bank 2

 

 

 

 

 

 

 

 

 

 

 

For orientation, the D-sub connector end of the module is facing you.

 

 

 

 

 

 

 

COM COM

 

 

 

 

 

 

 

 

 

1

 

2

3

4

5

6

1H

1L

7

8

 

27

28

9

10

39

40

Interlock1

1

 

2

3

4

5

6

7

8

9

10

 

11

12

13

14

15

16

17

Reserved

21

22

13

14

 

33

34

25

26

17

18

37

38

29

30

Interlock 1

18

 

19

20

21

22

 

23

24

25

26

27

28

29

30

31

32

 

33

GND

11

12

31

32

23

24

15

16

35

 

36

19

20

 

Reserved

 

 

 

 

 

 

34

 

35

36

37

38

39

40

41

42

43

 

44

45

46

47

48

49

50

WARNING As a safety feature, interlock 1 pins (17 and 33) on Bank 1 must be shorted to enable the Bank 1 Analog Bus relays to close. The optional 34923T-002 (for

1-wire) shorts these pins for you. This feature protects inadvertent routing of high voltages from the Analog Bus to the D-sub connector of the module.

Description

Pin

Description

Pin

Description

Pin

Description

Pin

Description

Pin

 

 

 

 

 

 

 

 

 

 

1

1

11

35

21

19

31

37

COM1 H

7

 

 

 

 

 

 

 

 

 

 

2

2

12

36

22

20

31

38

COM1 L

8

 

 

 

 

 

 

 

 

 

 

3

3

13

21

23

39

33

23

Interlock 1

17

 

 

 

 

 

 

 

 

 

 

4

4

14

22

24

40

34

24

Interlock 1

33

 

 

 

 

 

 

 

 

 

 

5

5

15

41

25

25

35

43

GND

34

 

 

 

 

 

 

 

 

 

 

6

6

16

42

26

26

36

44

Reserved

18

 

 

 

 

 

 

 

 

 

 

7

9

17

27

27

11

37

29

Reserved

47

 

 

 

 

 

 

 

 

 

 

8

10

18

28

28

12

38

30

Reserved

48

 

 

 

 

 

 

 

 

 

 

9

13

19

45

29

31

39

15

Reserved

49

 

 

 

 

 

 

 

 

 

 

10

14

20

46

30

32

40

16

Reserved

50

 

 

 

 

 

 

 

 

 

 

Bank 2

 

 

 

 

 

 

COM COM

 

 

 

 

 

 

 

 

 

 

41

42

43

44

45

46

2H

 

2L

47

48

 

67

68

49

50

79

80 Interlock 2

1

2

3

4

5

6

7

 

8

9

10

 

11

12

13

14

15

16

17

Reserved 61

62

53

54

 

73

74

65

 

66

57

58

77

78

69

70

Interlock 2

18

19

20

21

22

 

23

24

25

 

26

27

28

29

30

31

32

 

33

GND

51

52

71

72

63

64

 

55

56

75

 

76

59

60

 

Reserved

 

 

 

 

 

 

 

34

35

36

37

38

39

40

 

41

42

43

 

44

45

46

47

48

49

50

WARNING As a safety feature, interlock 2 pins (17 and 33) on Bank 2 must be shorted to enable the Bank 2 Analog Bus relays to close. The optional 34923T-002 (for

1-wire) shorts these pins for you. This feature protects inadvertent routing of high voltages from the Analog Buses to the D-sub connector of the module.

Description

Pin

Description

Pin

Description

Pin

Description

Pin

Description

Pin

41

1

51

35

61

19

71

37

COM2 H

7

 

 

 

 

 

 

 

 

 

 

42

2

52

36

62

20

72

38

COM2 L

8

 

 

 

 

 

 

 

 

 

 

43

3

53

21

63

39

73

23

Interlock 2

17

 

 

 

 

 

 

 

 

 

 

44

4

54

22

64

40

74

24

Interlock 2

33

 

 

 

 

 

 

 

 

 

 

45

5

55

41

65

25

75

43

GND

34

 

 

 

 

 

 

 

 

 

 

46

6

56

42

66

26

76

44

Reserved

18

 

 

 

 

 

 

 

 

 

 

47

9

57

27

67

11

77

29

Reserved

47

 

 

 

 

 

 

 

 

 

 

48

10

58

28

68

12

78

30

Reserved

48

 

 

 

 

 

 

 

 

 

 

49

13

59

45

69

31

79

15

Reserved

49

 

 

 

 

 

 

 

 

 

 

50

14

60

46

70

32

80

16

Reserved

50

 

 

 

 

 

 

 

 

 

 

34980A User’s Guide

119

Page 131
Image 131
Agilent Technologies Switch/Measure manual 34923A D-Sub Connectors for One-Wire Mode