Agilent Technologies Switch/Measure manual Sweep Count, Sweep count

Models: Switch/Measure

1 216
Download 216 pages 59.53 Kb
Page 62
Image 62

2Features and Functions

Sweep Count

The sweep count sets the number of sweeps per trigger event during a scan (a sweep is one pass through the scan list). The front- panel sample annunciator (“ * ”) turns on during each measurement.

Trigger

Sweep 1 Sweep 2

Sweep n

Trigger

. . .

t

Sweep Count

(1 to 500,000 sweeps)

Sweep count

The sweep count is valid only while scanning. If no channels have been assigned to the scan list, the specified sweep count is ignored (no error is generated).

You can specify a sweep count in conjunction with a trigger count and a sample count. The three parameters operate independent of one another, and the total number of readings returned will be the product of the three parameters.

You can store at least 500,000 readings in memory and all readings are automatically time stamped. If memory overflows, the new readings will overwrite the first (oldest) readings stored; the most recent readings are always preserved.

The CONFigure and MEASure? commands automatically set the sweep count to 1 sweep.

The instrument sets the sweep count to 1 after a Factory Reset

(*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.

Front Panel Operation: Scan (Configure) > SWEEP COUNT

Remote Interface Operation: SWEep:COUNt

50

34980A User’s Guide

Page 62
Image 62
Agilent Technologies Switch/Measure manual Sweep count, Front Panel Operation Scan Configure Sweep Count