The 34980A provides four 2- wire internal Analog Buses for easier signal routing. You can route your measurements directly to the internal DMM using the 34980A multiplexer and matrix modules, or you can connect to external signals via the Analog Bus connector located on the instrument’s rear panel (see connector pinout below). Since four 2- wire buses are provided, you can dedicate one bus for use with the internal DMM and use the other three buses for module extensions or additional signal routing between modules.

ANALOG

BUSSES

ABus1 HI (Pin 9)

ABus2 HI (Pin 8)

ABus3 HI (Pin 7)

ABus4 HI (Pin 6)

9 5

6 1

Current I (Pin 5)

ABus1 LO (Pin 4)

ABus2 LO (Pin 3)

ABus3 LO (Pin 2)

ABus4 LO (Pin 1)

Analog Bus connector (as viewed from rear of instrument)