Agilent Technologies Switch/Measure manual Automatic Trigger Delays

Models: Switch/Measure

1 216
Download 216 pages 59.53 Kb
Page 35
Image 35

Features and Functions

2

Automatic Trigger Delays

If you do not specify a trigger delay, the instrument selects a delay

for you. The delay is determined by the function, range, integration time, and ac filter setting as shown below.

DC Voltage, Thermocouple, DC Current (for all ranges):

Integration Time

Trigger Delay

 

 

PLC > 1

2.0 ms

PLC 1

1.0 ms

 

 

Resistance, RTD, Thermistor (2- and 4-wire):

 

Trigger Delay

Range

(for PLC > 1)

 

 

100

2.0 ms

1 k

2.0 ms

10 k

2.0 ms

100 k

25 ms

1 M

30 ms

10 M

200 ms

100 M

200 ms

 

 

 

Trigger Delay

Range

(for PLC 1)

 

 

100

1.0 ms

1 k

1.0 ms

10 k

1.0 ms

100 k

20 ms

1 M

25 ms

10 M

200 ms

100 M

200 ms

 

 

AC Voltage, AC Current (for all ranges):

AC Filter

Trigger Delay

 

 

Slow (3 Hz)

7.0 seconds

Medium (20 Hz)

1.0 second

Fast (200 Hz)

120 ms

 

 

Frequency, Period:

AC Filter

Trigger Delay

 

 

Slow (3 Hz)

600 ms

Medium (20 Hz)

300 ms

Fast (200 Hz)

100 ms

 

 

Digital Input, Totalize:

Trigger Delay

0 seconds

34980A User’s Guide

23

Page 35
Image 35
Agilent Technologies Switch/Measure manual Automatic Trigger Delays, Resistance, RTD, Thermistor 2- and 4-wire