Matrix Switch Modules

5

34933T-001 Terminal Block for Two-Wire Mode

This terminal block with screw- type connections is labeled with the model number and the abbreviated module name. In addition, space is available on the label for you to write the slot number.

N O TE

All modules that connect to the internal DMM are interlock

protected. This means that when an installed module is exposed (no

 

 

terminal block or cable is connected), the Analog Bus relays, which

 

are on Matrix 2, are open and disconnected from the Analog Buses.

 

See page 138 for further information.

 

 

The 34980A Product Reference CD (shipped with the instrument) contains a the 34933T (2- wire) Wiring Log for you to document your wiring configuration for this module. You can open the wiring log file in Microsoft® Excel® or Adobe® Acrobat® format

N O TE

If you are using an Agilent terminal block to connect your DUT to

this module be sure to use the 34933T-001 terminal block that

 

 

corresponds to the 2-wire configuration mode. Note that an error

 

will not be generated if you have installed a terminal block that

 

doesn't match the present module configuration.

 

 

34980A User’s Guide

157

Page 169
Image 169
Agilent Technologies Switch/Measure manual 34933T-001 Terminal Block for Two-Wire Mode