Agilent Technologies Switch/Measure manual Automatic Channel Delays

Models: Switch/Measure

1 216
Download 216 pages 59.53 Kb
Page 66
Image 66

2Features and Functions

To ensure you are getting the most accurate measurements possible, use care when setting the channel delay less than the default value (automatic). The default channel delay is designed to optimize parameters, such as settling time, for the most accurate measurements.

The CONFigure and MEASure? commands set the channel delay to automatic. A Factory Reset (*RST command) also sets the channel delay to automatic.

Front Panel Operation: Channel (Configure) > CHANNEL DELAY > TIME

Once you have added the specified channel to the scan list, the channel delay choice will be visible in the menu.

Interface Operation: The following command add a 2- second channel delay to the specified channels.

ROUT:CHAN:DELAY 2,(@1003,1013)

Automatic Channel Delays

If you do not specify a channel delay, the instrument selects a delay for you. The delay is determined by the delay based on function, range, integration time, and ac filter setting.

DC Voltage, Thermocouple, DC Current (for all ranges):

Integration Time

Channel Delay

 

 

PLC > 1

2.0 ms

PLC 1

1.0 ms

 

 

Resistance, RTD, Thermistor (2- and 4-wire):

 

Channel Delay

Range

(for PLC > 1)

 

 

100

2.0 ms

1 k

2.0 ms

10 k

2.0 ms

100 k

25 ms

1 M

30 ms

10 M

200 ms

100 M

200 ms

 

 

 

Channel Delay

Range

(for PLC 1)

 

 

100

1.0 ms

1 k

1.0 ms

10 k

1.0 ms

100 k

20 ms

1 M

25 ms

10 M

200 ms

100 M

200 ms

 

 

54

34980A User’s Guide