5Matrix Switch Modules

34932T Terminal Block

This terminal block with screw- type connections is labeled with the model number and the abbreviated module name. In addition, space is available on the label for you to write the slot number.

N O TE

All modules that connect to the internal DMM are interlock

protected. This means that when an installed module is exposed (no

 

 

terminal block or cable is connected), the Analog Bus relays, which

 

are on Matrix 2, are open and disconnected from the Analog Buses.

 

See page 138 for further information.

 

 

The 34980A Product Reference CD (shipped with the instrument) contains a the 34932T Wiring Log for you to document your wiring configuration for this module. You can open the wiring log file in Microsoft® Excel® or Adobe® Acrobat® format

Although they have separate screw-type connectors, rows labeled the same on a matrix are electrically connected. Therefore, you can wire the same-matrix rows in two locations.

Although columns are numbered the same on the two matrices, they are electrically separate. (example, C13)