Agilent Technologies Switch/Measure manual 34933A D-Sub Connectors for Two-Wire Mode, Matrix

Models: Switch/Measure

1 216
Download 216 pages 59.53 Kb
Page 168
Image 168

5Matrix Switch Modules

34933A D-Sub Connectors for Two-Wire Mode

Matrix 1

 

 

 

 

 

 

Matrix 1

 

Matrix 2

 

 

 

 

 

 

 

 

 

 

 

For orientation, the D-sub connector end of the module is facing you.

NOTE:

C4H

C4L

C4H

C4L

R4H

R4L

C5H

C5L

C5H

 

C5L

NC

NC

C7H

C7L

 

 

C7H

C7L

NC

bypass bypass

bypass

bypass

 

bypass bypass

1

2

3

 

4

5

6

7

8

9

 

10

11

12

13

14

 

 

15

16

17

 

C3H

C3L

C1H

C1L

 

 

C6H

C6L

 

 

 

 

 

 

 

 

 

NC

bypass bypass

bypass bypass C2H

C2L bypass bypass R2H

R2L

 

C8H

C8L

NC

NC

GND

 

18

 

19

20

21

22

 

23

24

25

26

27

28

 

29

30

31

 

32

33

 

NC

C3H

C3L

C1H

C1L

R3H

R3L

C6H

C6L

 

C2H

C2L

NC

NC

C8H

 

 

C8L

R1H

R1L

 

bypass bypass

bypass

bypass

34

35

36

 

37

38

39

40

41

42

 

43

44

45

46

47

 

 

48

49

50

In this diagram and the table below, R represents “row,” and C represents “column.”

Bypass” means to bypass the 100in-rush resistor

that protects the reed relays.

Description

Pin

Description

Pin

Description

Pin

Description

Pin

Description

Pin

Description

 

Pin

R1H

49

C1H

37

C5H

7

C1H bypass

21

C5H bypass

9

GND

 

33

 

 

 

 

 

 

 

 

 

 

 

 

R1L

50

C1L

38

C5L

8

C1L bypass

22

C5L bypass

10

No Connect

pins:

 

 

 

 

 

 

 

 

 

 

11-12, 17-18, 31-32,

R2H

27

C2H

23

C6H

41

C2H bypass

43

C6H bypass

25

 

 

 

 

 

 

 

 

 

 

34, and 45-46

 

R2L

28

C2L

24

C6L

42

C2L bypass

44

C6L bypass

26

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R3H

39

C3H

35

C7H

13

C3H bypass

19

C7H bypass

15

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R3L

40

C3L

36

C7L

14

C3L bypass

20

C7L bypass

16

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R4H

5

C4H

1

C8H

29

C4H bypass

3

C8H bypass

47

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R4L

6

C4L

2

C8L

30

C4L bypass

4

C8L bypass

48

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Matrix 2

 

 

C4H

C4L

 

 

C5H

C5L

C5H

 

C5L

 

 

 

 

 

 

C7H

C7L

 

C4H

C4L bypass bypass

R8H

R8L

bypass

bypass

NC

NC

C7H

C7L

 

bypass bypass

Interlock

1

2

3

4

5

6

7

8

9

 

10

11

12

13

14

 

 

15

16

17

 

C3H

C3L

C1H

C1L

 

 

C6H

C6L

 

 

 

 

 

 

 

 

 

NC

bypass bypass bypass bypass C2H

C2L

bypass bypass R6H

R6L

 

C8H

C8L

NC

NC

Interlock

18

19

20

21

22

23

24

 

25

26

27

28

 

29

30

31

 

32

 

33

NC

C3H

C3L

C1H

C1L

R7H

R7L

C6H

C6L

 

C2H

C2L

NC

NC

C8H

 

 

C8L

R5H

R5L

 

bypass bypass

bypass

bypass

34

35

36

37

38

39

40

41

42

 

43

44

45

46

47

 

 

48

49

50

NOTE:

In this diagram and the table below, R represents “row,” and C represents “column.”

“Bypass” means to bypass the 100in-rush

resistor that protects the reed relays.

WARNING As a safety feature, interlock pins (17 and 33) must be shorted to enable the Analog Bus relays, which are on Matrix 2, to close. The optional 34933T-001 (for 2-wire) terminal block shorts these pins for you. This feature protects inadvertent routing of high voltages from the Analog Bus to the D-sub connector of the module.

Description

Pin

Description

Pin

Description

Pin

Description

Pin

Description

Pin

R5H

49

C2H

23

C7H

13

C4H bypass

3

Interlock

17

 

 

 

 

 

 

 

 

 

 

R5L

50

C2L

24

C7L

14

C4L bypass

4

Interlock

33

 

 

 

 

 

 

 

 

 

 

R6H

27

C3H

35

C8H

29

C5H bypass

9

No Connect

pins:

 

 

 

 

 

 

 

 

11-12, 18, 31-32, 34,

R6L

28

C3L

36

C8L

30

C5L bypass

10

 

 

 

 

 

 

 

 

and 45-46

 

R7H

39

C4H

1

C1H bypass

21

C6H bypass

25

 

 

 

 

 

 

 

 

 

 

 

 

 

R7L

40

C4L

2

C1L bypass

22

C6L bypass

26

 

 

 

 

 

 

 

 

 

 

 

 

R8H

5

C5H

7

C2H bypass

43

C7H bypass

15

 

 

 

 

 

 

 

 

 

 

 

 

R8L

6

C5L

8

C2L bypass

44

C7L bypass

16

 

 

 

 

 

 

 

 

 

 

 

 

C1H

37

C6H

41

C3H bypass

19

C8H bypass

47

 

 

 

 

 

 

 

 

 

 

 

 

C1L

38

C6L

42

C3L bypass

20

C8L bypass

48

 

 

 

 

 

 

 

 

 

 

 

 

156

34980A User’s Guide

Page 168
Image 168
Agilent Technologies Switch/Measure manual 34933A D-Sub Connectors for Two-Wire Mode, Matrix