Agilent Technologies Switch/Measure manual DMM or Channel Configure Scaling Gainoffsetunits

Models: Switch/Measure

1 216
Download 216 pages 59.53 Kb
Page 53
Image 53

Features and Functions

2

The CONFigure and MEASure? commands automatically set the gain (“M”) to 1 and offset (“B”) to 0.

A Factory Reset (*RST command) turns off scaling and clears the scaling values on all channels. An Instrument Preset (SYSTem:PRESet command) does not clear the scaling values and does not turn off scaling.

Front Panel Operation:

DMM or Channel (Configure) > SCALING > GAINOFFSETUNITS

To define the label on the selected channel, press the arrow keys to move the cursor to a specific position and then turn the knob to select the desired letter or number. To clear the label on the selected channel, change each character to “ ^ ” (starting with the rightmost character) and then press the left arrow key to move to the next character.

Remote Interface Operation: Use the following commands to set the gain, offset, and custom measurement label.

CALC:SCALE:GAIN 1.2,(@1003)

CALC:SCALE:OFFSET 10,(@1003)

CALC:SCALE:UNIT 'PSI',(@1003)

After setting the gain and offset values, send the following command to enable the scaling function on the specified channel.

CALC:SCALE:STATE ON,(@1003)

34980A User’s Guide

41

Page 53
Image 53
Agilent Technologies Switch/Measure manual DMM or Channel Configure Scaling Gainoffsetunits