Features and Functions

2

AC Voltage, AC Current (for all ranges):

AC Filter

Channel Delay

 

 

Slow (3 Hz)

7.0 seconds

Medium (20 Hz)

1.0 second

Fast (200 Hz)

120 ms

 

 

Frequency, Period:

AC Filter

Channel Delay

 

 

Slow (3 Hz)

600 ms

Medium (20 Hz)

300 ms

Fast (200 Hz)

100 ms

 

 

Digital Input, Totalize:

Channel Delay

0 seconds

Front Panel Operation: Channel (Configure) > CHANNEL DELAY > AUTO

Once you have added the specified channel to the scan list, the channel delay choice will be visible in the menu.

Interface Operation: The following command enables an automatic channel delay on the specified channels.

ROUT:CHAN:DELAY:AUTO ON,(@1003,1013)

Selecting a specific channel delay using the ROUTe:CHANnel:DELay

command (see “Channel Delay” on page 53) disables the automatic

channel delay.