Features and Functions

2

To initiate a scan and store all readings in memory, press

Scan (Measure). Each time you initiate a new scan, the instrument clears all previously stored readings. If you have not defined a scan list,

Scan (Measure) performs an internal DMM scan independent of any channels.

To stop a scan in progress, press and hold Scan (Measure).

To Build a Scan List From the Remote Interface

Use the ROUTe:SCAN command to define the list of channels in the scan list. To determine what channels are currently in the scan list, use the ROUTe:SCAN? query command.

To add channels to the present scan list, use the ROUTe:SCAN:ADD command. To remove channels from the present scan list, use the

ROUTe:SCAN:REMove command.

To remove all channels from the scan list, send “ROUT:SCAN (@)”.

To initiate a scan, use the INITiate or READ? command. Measurements are stored in memory. Each time you initiate a new scan, the instrument will clear the previous set of readings from memory.

To stop a scan in progress, use the ABORt command.

34980A User’s Guide