Instruction Manual

IB-106-300NH Rev. 4.2 July 2002

World Class 3000

Table 5-5. SETUP Sub-Menu

SUB-MENU

 

 

 

SELECTION

PARAMETERS

DESCRIPTION

Calibration

High Gas

____% O2

Value of high O2 calibration gas (0.1000% - 25.00%

 

 

 

O2).

Low Gas

____% O2

Auto Cal

YES/NO

Output Tracks

YES/NO

Cal Intrvl

XD XH

Next Cal

XH

Value of low O2 calibration gas (0.1000% - 25.00% O2).

MPS required for Auto Cal.

NO, locks output during calibration.

Select time between calibrations in number of days and hours (1 year max).

Time until next calibration in number of hours (1 year max).

 

Gas Time

0:30 - 20:00

 

 

 

 

 

 

Amount of time calibration gases will be turned on in

 

 

 

 

 

 

 

 

 

 

 

 

number of minutes and seconds; allow enough time

 

 

 

 

 

 

 

 

 

 

 

 

for signal values to stabilize.

 

Purge Time

0:30 - 20:00

 

 

 

 

 

 

Amount of time for gas lines to clear of calibration

 

 

 

 

 

 

 

 

 

 

 

 

gas.

 

Res Alarm

50 W – 10 kW

 

 

 

Resistance alarm set from 50 to 10,000 ohms.

O2 Calculation

Slope

____ mV/D

 

 

 

Set value between 34.5 and 57.5.

 

Constant

____ mV

 

 

 

Set value between -20.0 and +20.0 mV.

 

Set Point

____°C

 

 

 

Set either 736 for World Class 3000 probes or 843

 

 

 

 

 

 

 

 

 

 

 

 

for 218 probes.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ensure the correct voltage is selected when using HPS 3000 with either World Class 3000 probes or 218 probes. Refer to Figure 2-15,Jumper Selection Label for proper voltage selections. If incorrect SET POINT is selected, damage to the probe may occur.

 

Reset slope and constants.

Press ENTER to reset slope and constants to values

 

 

 

from the latest successful calibration.

O2 Alarms

Hi Alarm

____% O2

Set value for high alarm limit (0.1000% - 25.00%).

 

Lo Alarm

____% O2

Set value for low alarm limit (0.1000% - 25.00%).

 

Alarm DB

____% O2

Set value for alarm dead band (0.0000% - 25.00%).

Efficiency Calc.

Enable Calc.

YES/NO

Select YES to enable, NO to disable.

 

K1 Value

_______

Set between 0.0000 and 1.000. Refer to Table 5-6.

 

K2 Value

_______

Set between 0.0000 and 1.000. Refer to Table 5-6.

 

K3 Value

_______

Set between 1.000 and 20.00. Refer to Table 5-6.

 

 

 

 

 

5-12 General User Interface (GUI) Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 82