Instruction Manual

IB-106-300NH Rev. 4.2 July 2002

World Class 3000

Table 6-3. Cell Troubleshooting (continued)

Problem

Cause

Corrective Action

Status is ResHi or CalEr.

Cell mV = -20 to 120 mV (normal) (continued).

1. Reference air contamination (oil/water).

Clean or replace lines and valves as needed.

2. Cell leads reversed.

Check cell signal wiring from probe junction box to electronics, and correct wiring if needed.

3.Reference/test gas lines reversed. Switch piping as needed.

4.Diffusion element fault. Diffusion element cracked, broken, missing, or plugged.

Replace diffusor or snubber as needed. Diffusors are disposable because it is difficult to clean a diffusor and know the tiny pores are open. A flow and pressure test with a manometer is possi- ble but usually not practical. To clean a snubber, blow off surface dirt with pressurized air and clean the unit in an ultrasonic bath.

5. Faulty cell. Low sensor cell output when test gas is applied.

If test gas flow is good and there is low cell signal, replace the cell, or call the SCAN line for assistance.

Typical cell output:

Test Gas

mV

8.0%

18 to 25

0.4%

76 to 86

6. Cell performance degraded from aging.

Replace the sensor cell if its resistance has increased beyond 1 kOhm and the slope calculated during calibration has decreased lower than 40 mV/decade.

7. Electronics fault. Cell output is good, and the input to the electronics is good.

Check the electronics package. In an IFT, replace the microprocessor or interface board as needed. In a CRE, replace the DPI board if needed.

Status is Res Hi.

Cell mV = -120 to 20 mV.

1. Cell leads reversed.

Check cell signal wiring from probe junction box to electronics, and correct wiring if needed.

2.Reference/test gas lines reversed. Switch piping as needed.

3.Reference air (nitrogen).

Confirm labels on test gas bottles are correct. 100% nitrogen must NOT be used as a zero gas because cell protection will engage and affect the O2 reading. Reference air should be clean, dry instrument air prepared from ambient air with 20.95% O2.

6-6 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 90
Image 90
Emerson B-106-300NH instruction manual Status is Res Hi Cell mV = -120 to 20 mV