Instruction Manual
World Class 3000
|
| Table |
|
| |
Off | Heater power has been turned OFF by the electronics. The display shows 0% O2. Several con- | |
| ditions may cause the OFF status: | |
| 1. | The cell heater temperature is below |
| 2. | The cell temperature is more than 70°C above the set point. The heater is out of control. |
|
| The triac module may have failed. |
| 3. | The cell heater thermocouple voltage has remained within +1.5 mV for more that 4 minutes. |
|
| The thermocouple may be shorted. |
| 4. | The AD590 voltage is below 50.0 mV (50K or |
| 5. | The AD590 voltage is above 363 mV (363K or 90°C). If HPS is used with IFT, then IFT in- |
|
| terconnect board has JM1 in position connecting two AD590s in parallel. |
PrbEr | The probe is disconnected or cold, or leads are reversed. | |
|
| |
HtrEr | There is a fault within the heater system. The heater temperature is more than +25°C from the | |
| set point. When the unit is first turned ON, HtrEr is normal. The heater may take 0.5 to 1.0 | |
| hours to warm up. | |
InCal | The system is currently undergoing calibration. If Output Tracks is set to YES, the output will | |
| show changing O2 values. If Output Tracks is set to NO, the output will hold the | |
| value. | |
LowO2 | The measured O2 value is below the low O2 alarm limit. The problem may be in the probe or the | |
| process. | |
HiO2 | The measured O2 value is above the high O2 alarm limit. The problem may be in the probe or | |
| the process. | |
NoGas | Test gas pressure is too low. Pressure switches are set to trigger this alarm at 12 to 16 psig (83 | |
| to 110 kPa gage). Test gas regulators are usually set at 20 to 25 psig (138 to 172 kPa gage). | |
| Possible causes are: | |
| 1. | At least one test gas pressure switch is open. |
| 2. | A test gas cylinder is empty. |
| 3. | There is an MPS or piping failure. |
| 4. | If MPS is not connected, CALRET and NOGAS signals should be jumpered on the intercon- |
|
| nect board. |
CalEr | An error occurred during the last calibration. The error may be one of the following: | |
| 1. | The new calculated slope value is outside the range 34.5 to 57.6 mV/decade. |
| 2. | The new calculated constant value is outside the range +20.0 to |
| 3. | The test gas pressure switch opened during calibration. |
| Ensure that the proper test gases are being used, and that the gas flows are set properly. Refer | |
| to Appendix D for additional MPS troubleshooting information. | |
ResHi | The resistance calculated during the last good calibration was greater than the High Resistance | |
| Alarm limit set in the calibration setup. The resistance limit may be set wrong, or there is a | |
| problem with the probe. | |
Ok | Operation appears to be normal. | |
|
| |
(blank | A possible failure within the IFT. Check the LED on the microprocessor board to help isolate | |
screen) | problems. See IFT Problem in the troubleshooting tables. | |
|
|
|
Rosemount Analytical Inc. A Division of Emerson Process Management |