CALIBRATION GAS KIT #1 (P/N 6296A27G01)

27270007

Figure 4-2. Portable Rosemount Oxygen

Calibration Gas Kit

2Extra gas bottles are available at:

Rosemount Analytical Inc. Box 901

Orrville, Ohio 44667 U.S.A.

Rosemount Limited Burymead Road Hitchin, Herts. U.K.

Rosemount Italy

VIA Guido Cavalcanti 8 20127 Milan, Italy

Rosemount Spain Saturnino Calleja 6 28002 Madrid Spain

Rosemount P/N 3530B07G01 for probe 0.4% oxygen in ni- trogen in disposable bottle.

Rosemount P/N 3530B07G02 for probe 8% oxygen in nitro- gen in disposable bottle.

3A check valve is required at the probe (between the cali- bration fitting and the gas line) to prevent the migration of process gases down the cali-

bration gas line.

4

(b)Calibration with a Portable Rose- mount Oxygen Calibration Gases Kit.

1A typical portable calibration setup is shown in Figure 4-3.For manual (semiautomatic) calibration, remove cap plug from the calibrate in place fit- ting. The cap plug must be retained to seal this fitting af- ter calibration is complete; failure to do so may render the probe useless if the sys- tem pressure is slightly nega- tive. The reference air should be flowing as in normal operation.

2Refer to paragraph 4-2.d of this section for Manual (Semi- automatic) Calibration setup and procedure using the IFT.

3Screw the pushbutton regu- lator with contents gage on to the calibration gas of choice and inject the calibration gas by opening the valve. Gas is on continuously when the valve is opened.