Emerson B-106-300NH instruction manual Figure B-6. HPS Troubleshooting Flowchart, #3

Models: B-106-300NH

1 182
Download 182 pages 23.73 Kb
Page 131
Image 131

World Class 3000

Instruction Manual

Appendix B Rev. 2.2 January 1997

SYMPTOM

PROBE HEATER START TO HEAT UP AND THEN LOOSES TEMPERATURE. (MODEL TC200 ONLY)

CHECK PARAMETER 35 ON TC200. IF THE NUMBER IS NEGATIVE THEN SOMEWHERE THE THERMOCOUPLE WIRES ARE REVERSED.

B

35730005

Figure B-6. HPS Troubleshooting Flowchart, #3

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendices B-7

Page 131
Image 131
Emerson B-106-300NH instruction manual Figure B-6. HPS Troubleshooting Flowchart, #3