Instruction Manual

Appendix D Rev. 2.4 July 1998

MPS 3000 TROUBLESHOOTING

D-3 OVERVIEW

This section describes troubleshooting for the Multiprobe Calibration Gas Sequencer. Addi- tional troubleshooting information can be found in the Instruction Bulletin for the electronics package.

Install all protective equipment covers and safety ground leads after trouble- shooting. Failure to replace covers and ground leads could result in seri- ous injury or death.

D-4 TROUBLESHOOTING

Table D-2provides a guide to fault finding fail- ures within the MPS. The flowchart in Figure

D-5provides an alternate approach to faultD finding MPS related problems.

Table D-2. Fault Finding

 

SYMPTOM

CHECK

FAULT

REMEDY

 

 

 

 

 

1.

Power to solenoid,

Calibration gas

Insufficient calibration

Install new calibration gas

 

calibration gas not

 

gas

tanks.

 

released to probe.

 

 

 

 

 

Solenoid

Solenoid failure

Replace solenoid.

2.

No power to

Power supply output

Power supply failure

Replace power supply.

 

solenoid.

 

 

 

 

 

Fuses in power

Fuse blown

Replace fuse.

 

 

supply

 

 

 

 

Main power source

Main power off

Reestablish power.

 

 

 

 

 

Rosemount Analytical Inc. A Division of Emerson Process Management