3.6.6 Calibrating an analog input

To calibrate an analog input, do the following:

1.Select Analog Input... from the Hardware menu. The Analog Input window displays.

Figure 3-29.The Analog Inputs window

2.Click on the analog input that you want to calibrate.

3.Set the analog input’s Zero Scale by entering its minimum anticipated value.

4.Set the analog input’s Full Scale by entering your maximum anticipated value.

5.Click AutoCal...(F4) or press F4. The Analog Input Calibration Wizard runs.