MON20/20 Software for Gas Chromatographs

 

User Manual

 

OCTOBER 2010

3-9000-745

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 2-4.The Chromatogram Viewer

 

 

 

 

 

 

 

 

 

 

timed events

 

 

chromatogram

 

 

 

 

 

 

 

window

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

component

 

 

 

 

 

data

Note

By default, the timed events and component data tables are configured to scroll to and highlight the next occurring event in the analysis cycle. To disable this feature, right- click on one of the tables and uncheck the Auto Scroll option on the pop-up menu.

2.1.3 Viewing an archived chromatogram

Archived chromatograms are stored on the GC, so you must be logged in to access them. With MON 20/20 archived chromatograms as old as four days are available for viewing.

Archived chromatograms are sorted and displayed on four tabbed panes:

Chromatograms - This view displays all chromatogram types sorted by time so that the newest file is always listed first. This view can be further configured to display only the files from the last five runs for each stream, or to display all the files that are stored on the GC.