User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
4-98
4.15 Mapping Foundation Fieldbus variables
To map a GC variable to a Foundation Fieldbus process variable (PV), do
the following:
1. Select FFB PV Mappings... from the Application menu. The FFB
PV Mappings window displays.
Figure 4-86. The FFB PV Mappings window
2. Select a new variable by clicking on the appropriate drop-down list
under the Variable column. For a demonstration of how to use the
context-sensitive variable selector, see “Using the context-sensitive
variable selector” on page 1-42.
Note
The PV Value column displays the current value of the GC variable indicated in the
Variable column.