User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 OCTOBER 2010
1-36
1.6 Restoring the GC to its factor y settings
The GC’s default timed event, component data and validation data tables
are created at the factory and are not accessable by users. To restore
these tables to their default values, do the following:
Note
The GC should be in Idle mode while performing this task.
1. With the GC idle, select Restore to Factory Settings... from the
File menu. The following warning message displays:
Figure 1-22. Restore to Factory Settings warning message
2. Click Yes. The MON 20/20 restores the default values to the GC’s
data tables. When the process is completed, the following message
displays:
Figure 1-23. Restoration completed message
3. Click OK.