User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 OCTOBER 2010
B-4
B.3.1 Concentration analysis with response factor
Calibration
The concentration calculations discussed as follows require a unique
response factor for each component in an analysis. These factors may be
manually entered by an operator or automatically calculated by
calibrating the system.
where
Calculated response factors are stored by the GC for use in the
concentration calculations, and are printed out in the configuration and
calibration reports.
or
ARFnArea response factor for component n in area per mole percent (%).
HRFnHeight response factor for component n.
AreanArea associated with component n in calibration gas.
HtnHeight associated with component n in mole percent in calibration gas.
CalnAmount of component n in mole percent of calibration gas.
ARFn
Arean
Caln
---------------
=HRFn
Htn
Caln
-----------
=
RFAVGn
RFi
i1=
k
k
------------------
=