User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 OCTOBER 2010
B-12
Note
All components in the sample must be measured in order to calculate LRDT.
where
Note
All components in the sample must be measured in order to calculate liquid density.
where
LRDTLiquid Relative Density of sample, relative to water at 60 °F.
LRDiLiquid Relative Density of component i (GPA2145-94).
LVpercent Liquid Volume Percent.
LDTLiquid Density of total sample in pounds per gallon
LDiLiquid Density of component i (GPA 2145-94).
LViLiquid Volume Percent of component i.
LRDT
LVpercenti
()LRDi
()
i1=
k
100
--------------------------------------------------------------
=
LDT
LVi
()LDi
()
i1=
k
100
--------------------------------------
=
GD RD()76.4976()=