Xi Advanced Electronics

Instruction Manual

IM-106-910Xi, Original Issue November 2010

SYSTEM

A typical system installation for a Xi and O2 Probe is shown in Figure 2-1.

CONSIDERATIONS

 

Figure 2-1. Typical System

Installation

MECHANICAL

INSTALLATION

Xi Advanced Electronics

The Xi Advanced Electronics is available in a panel mounting, wall mounting, or pipe mounting configuration. Refer to Figure 2-2 or Figure 2-3 for the panel, wall, or pipe mounting details.

1.Ensure all components are available to install the Xi.

2.Select a mounting location near or removed from the O2 Probe. Consider the temperature limitations of the Xi (see "Specifications") when selecting the mounting location.

3.Mount the Xi at a height convenient for viewing and operating the interface. Approximately 5 ft (1,5 m) is recommended.

4.The keypad window on the Xi may have interior and exterior protective membranes. Remove the protective membranes prior to use of the Xi enclosure. Failure to remove the protective membranes may cause the display to appear distorted. The membrane may be difficult or impossible to remove after extended use at elevated temperatures.

2-2

Page 22
Image 22
Emerson Process Management IM-106-910Xi System, Considerations, Mechanical Installation, Xi Advanced Electronics