Emerson Process Management IM-106-910Xi Electrical Installation, Xi Advanced Electronics

Models: IM-106-910Xi

1 126
Download 126 pages 27.6 Kb
Page 25
Image 25

Instruction Manual

IM-106-910Xi, Original Issue November 2010

Xi Advanced Electronics

ELECTRICAL

INSTALLATION

Xi Advanced Electronics

All wiring must conform to local and national codes. Multiple wiring diagrams are shown in this section. Always refer to the diagrams that apply to your transmitter configuration and disregard all other wiring diagrams.

Disconnect and lock out power before connecting the power supply.

Install all protective covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.

To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, connection to the main electrical power supply must be made through a circuit breaker (min 10A) which will disconnect all current-carrying conductors during a fault situation. This circuit breaker should also include a mechanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.

NOTE

Line voltage, signal, and relay wiring must be rated for at least 105° C (221° F).

NOTE

If metal conduit is used with the Xi the conduit should be reliably bonded to protective earth. The grounding plate inside the Xi is not bonded to PE and does not provide adequate grounding.

1.Remove cover screws from the front cover of the Xi. Swing down the front cover of the interface box.

2.Pull out the I/O board on the right-hand side of the card rack inside the Xi. If your system is configured to operate two transmitter probes there are two I/O interface boards.

3.See Figure 2-4. Connect the 4-20 mA signal wires at J4 of the I/O board. Attach the supplied ferrite clamp over the 4-20 mA OUT wires that extend past the shield.

NOTE

Installation of the ferrite clamp over the 4-20 mA OUT wires is required for compliance with the European EMC Directive.

2-5

Page 25
Image 25
Emerson Process Management IM-106-910Xi instruction manual Electrical Installation, Xi Advanced Electronics