Instruction Manual
IM-106-910Xi, Original Issue November 2010
Xi Advanced Electronics
ELECTRICAL
INSTALLATION
Xi Advanced Electronics
All wiring must conform to local and national codes. Multiple wiring diagrams are shown in this section. Always refer to the diagrams that apply to your transmitter configuration and disregard all other wiring diagrams.
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, connection to the main electrical power supply must be made through a circuit breaker (min 10A) which will disconnect all
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105° C (221° F).
NOTE
If metal conduit is used with the Xi the conduit should be reliably bonded to protective earth. The grounding plate inside the Xi is not bonded to PE and does not provide adequate grounding.
1.Remove cover screws from the front cover of the Xi. Swing down the front cover of the interface box.
2.Pull out the I/O board on the
3.See Figure
NOTE
Installation of the ferrite clamp over the