Xi Advanced Electronics

Instruction Manual

IM-106-910Xi, Original Issue November 2010

PROBE PARAMETER DESCRIPTIONS

PARAMETER

Among the parameters available through the Xi and 375/475 Field Communicator menus are a number of "Probe Parameters". The probe parameters define variables that configure a specific probe in the transmitter system. Probe parameters are described in the following table.

TX

I/O

NAME

UNITS

PARAMETER DESCRIPTION

Y

Y

O2

%

Current oxygen concentration value (O2%). The value should reflect the last good O2 value if it

 

 

 

 

is in the "Lock" state during calibration.

 

 

 

 

 

Y

Y

O2 Temp

degC

Current O2 sensor temperature.

Y

Y

CJC Temp

degC

Current cold junction temperature.

 

 

 

 

 

N

Y

Elec Temp

degC

Current electronic temperature measured at the I/O board.

Y

Y

O2 Cell

mV

Raw mV value for ZrO2 sensor.

N

Y

TC Volt

mV

O2 T/C voltage.

Y

Y

Cell Imp

Ohm

Cell impedance/sensor resistance measured.

 

 

 

 

 

Y

Y

Htr Volt

Volt

Heater voltage.

Y

Y

CPU Volt

Volt

Transmitter CPU voltage.

 

 

 

 

 

Y

Y

O2 AO

mA

Analog output value represents the O2 concentration measurement.

N

Y

O2 AO%

%

O2 analog output percentage for O2 AO.

 

 

 

 

 

Y

Y

O2 Temp Max

degC

This is the highest O2 sensor temperature reached since last reset.

Y

Y

CJC Temp Max

degC

This is the highest temperature reached at the cold junction since last reset.

 

 

 

 

 

N

Y

Elec Temp Max

degC

This is the highest temperature reached at the I/O board since last reset.

Y

Y

Htr Volt Max

degC

This is the highest heater voltage reached since last reset.

 

 

 

 

 

Y

Y

Htr Duty Cycle

--

O2 heater duty cycle. Value between 0 and 1.

Y

Y

PID SP

degC

PID temperature set point.

 

 

 

 

 

Y

Y

Htr Ramp Rate

degC/s

Heater ramp rate calculated in degree C per second.

N

Y

Flame Stat In

--

Flame status input state.

 

 

 

 

(OFF/ON)

 

 

 

 

 

N

Y

SPS/IMPS In

--

SPS/IMPS input state.

 

 

 

 

(OFF/ON)

N

Y

SPS/IMPS Out

--

SPS/IMPS output state.

 

 

 

 

(OFF/ON)

 

 

 

 

 

N

Y

Alm Relay 1 Out

--

Alarm Relay 1 output state.

 

 

 

 

(OFF/ON)

N

Y

Alm Relay 2 Out

--

Alarm Relay 2 output state.

 

 

 

 

(OFF/ON)

 

 

 

 

 

Y

Y

OP Mode

--

Device operating mode:

 

 

 

 

PO=Power up; WU=Warm Up (analog output is railed); NM=Normal operation;

 

 

 

 

CA=Calibrating (analog output can be tracking or locked at last good value based on

 

 

 

 

"AO Tracks" configuration); AL=Alarm detected (recoverable); SF=Alarm detected

 

 

 

 

(non-recoverable)

Y

Y

Tag

--

Device tag.

 

 

 

 

 

Y

Y

Device ID

--

Unique Device ID number. (HART)

Y

Y

PV is

--

Primary variable assignment. (HART)

 

 

 

 

 

Y

Y

SY is

--

Secondary variable assignment. (HART)

Y

Y

TY is

--

Third variable assignment. (HART)

 

 

 

 

 

Y

Y

QV is

--

Fourth variable assignment. (HART)

Y

Y

Cal Slope

mV/Dec

Current calibration slope. This is the slope value that was calculated as a result of the last

 

 

 

 

successful calibration.

 

 

 

 

 

Y

Y

Cal Const

mV

Current calibration constant. This is the constant value that was calculated as a result of the last

 

 

 

 

successful calibration. It is valid between -4mV and +10mV.

Y

Y

Cal Imp

Ohm

Cell Impedance. This is the sensor resistance that was calculated as a result of the last

 

 

 

 

successful calibration.

 

 

 

 

 

N

Y

Prev Slope

mV/Dec

Previous calibration slope. There are ten calibration results. 1 is the most recent and 10 is the

 

 

 

 

least recent calibration slope.

N

Y

Prev Const

mV

Previous calibration constant. There are ten calibration results. 1 is the most recent and 10 is

 

 

 

 

the least recent calibration constant.

4-8