Instruction Manual

IM-106-910Xi, Original Issue November 2010

Xi Advanced Electronics

Section 4 Startup and Operation

 

Overview

page 4-1

 

Startup

page 4-1

 

Operation via Xi

page 4-2

 

Startup Display

page 4-2

 

Error Conditions

page 4-2

 

Xi Controls

page 4-2

 

Password Protection

page 4-3

 

Xi Menu (Sheet 1 of 3)

page 4-4

 

System Parameter Descriptions

page 4-7

 

Probe Parameter Descriptions

page 4-8

 

Operation Via HART/AMS

page 4-10

 

Field Communicator Signal Line Connections

page 4-10

 

Field Communicator Menu Trees

page 4-11

 

Off-line and On-line Operations

page 4-11

 

Calibration - General

page 4-14

 

O2 Calibration

page 4-14

 

O2 Calibration with Xi

page 4-14

 

O2 Calibration with Xi and Field Communicator

page 4-15

 

D/A Trim

page 4-16

 

D/A Trim with Xi

page 4-16

OVERVIEW

 

Interface to the Xi for setup, calibration and diagnostics can be via a 375/475

 

Field Communicator or Asset Management System.

 

 

Setup, calibration and diagnostic operations will differ depending on the

 

selected interface for communications with the transmitter.

 

STARTUP

The O2 Probe will take approximately 45 minutes to warm up to the 736° C

 

heater setpoint. The 4-20 mA signal will remain at a default value of 3.5 mA

through this warm-up period. Once warm, the O2probe will be reading oxygen, and the 4-20 mA signal will be reading based on the default range of 0-10% O2.

NOTE

The Xi offers optional advanced features such as elevated process temperature capability to 800° C, autocalibration via an SPS solenoid box, a stoichiometer feature for indicating the level of oxygen deficiency in reducing conditions, and programmable reference to enhance accuracy at near ambient levels of O2.