Xi Advanced Electronics

Instruction Manual

IM-106-910Xi, Original Issue November 2010

Table 5-1. Diagnostic/Unit Alarm Fault Definitions

Message

Status

Self Clearing

NV Mem Fail

A checksum error was detected in the nonvolatile memory configuration data when the unit was

No

 

 

turned on.

 

 

 

 

Factory Mode

On Xi box I/O board, SW4, position 1 is set on "ON". On transmitter electronics board, SW1,

No

 

 

position 1 is set to "ON". This setting should only be used in the factory.

 

Board Temp Hi

The transmitter electronic board temperature reading is above 126° C or the

No

 

 

Xi unit I/O board temperature reading is above 86° C.

 

 

 

 

 

O2

Htr Ramp Rate

The O2 sensor heater ramp rate is greater than max allowed ramp rate indicating a run away

No

 

 

heater condition.

 

O2

Sensor Open

The cell impedance voltage is reading less than -1.1 VDC indicating the O2 sensor wires may be

Yes

 

 

disconnected or the O2 sensor junction may be open.

 

O2

TC Open

The O2 cell heater thermocouple voltage is reading more than the hardware configured threshold

Yes

 

 

voltage. This indicates the thermocouple wires may be disconnected or the thermocouple junction

 

 

 

may be open.

 

O2

Temp Low

The heater temperature is below the minimum temperature. The predefined low temperature

Yes

 

 

threshold is the 726° C.

 

 

 

 

 

O2

Temp High

The heater temperature is above the defined temperature threshold. The high temperature

Yes

 

 

threshold is the defined by the "High Temp Alm SP" parameter. The default value is 750° C.

 

O2

T/C Shorted

The O2 sensor heater temperature thermocouple voltage is shorted.

Yes

O2

T/C Reversed

The O2 sensor heater temperature thermocouple voltage is reading a negative voltage indicating

Yes

 

 

the thermocouple wire connections may be reversed.

 

O2

Htr Failure

The O2 sensor heater temperature is not responding to the controller and can't reach final

Yes

 

 

temperature set by the device, indicating the O2 heater may have failed.

 

No Flame

The Flame Status Relay Input is set to the OFF state indicating it is not safe to operate the O2

Yes

 

 

heater and the heater should be turned off.

 

Xmtr Disconnect

Communication failures detected between transmitter and the I/O board indicating the transmitter

Yes

 

 

has been disconnected from the I/O board.

 

 

 

 

Cal Recommended

Probe calibration is recommended. The cell impedance is above 100 ohms and has shifted 50

Yes

 

 

ohms since the last calibration; the accuracy of the O2 reading may be compromised.

 

Cal Failed

A calibration error occurred during the last calibration. The measured slope or constant is outside

Yes

 

 

the acceptable range. (Slope: 34.5 to 57.5 mv/decade) (Constant: ±20 mv)

 

 

 

 

Cell Imp High

The O2 sensor impedance/cell resistance value measurement is greater than 2000 Ohms

Yes

 

 

indicates the cell may be beyond its useful life.

 

CPU Voltage Low

The CPU voltage is less than 2.7 V indicating the CPU voltage is too low.

Yes

 

 

 

CPU Voltage High

The CPU voltage is more than 3.3 V indicating the CPU voltage is too high.

Yes

Htr Voltage Low

The heater voltage for the O2 cell heater is below 35 volts.

Yes

Htr Voltage High

The heater voltage for the O2 cell heater is above 264 volts.

Yes

Low O2

The O2 reading is below the "Low O2 Alm SP".

Yes

CALIBRATION PASSES, BUT STILL READS INCORRECTLY

There are a few fault conditions where no alarm indication is present and the probe passes calibration, but the O2 reading may still be incorrect:

An incorrect flow rate of calibration gases can cause a shifted calibration. If the flow rate of calibration gases is too low, process gases can mix in with the calibration gases causing a mixture at the cell that is different than what is noted on the calibration gas bottles. Always set the calibration flow rate when a new diffuser is installed, and never readjust this flow rate until another new diffuser is installed. For applications with heavy particulate loading, see "Probe Passes Calibration, But O2 Still Appears To Read Low".

No or improper reference air supplied.

5-4

Page 58
Image 58
Emerson Process Management IM-106-910Xi instruction manual Calibration PASSES, but Still Reads Incorrectly, Message Status