Xi Advanced Electronics
Instruction Manual
11.Reinstall wiring for Alarm Outputs, Flame Status Input and/or SPS/IMPS as applicable. See Figure
12.Connect the ribbon cable to the I/O Board. A new cable is supplied and should be used if the old one is damaged.
13.Swing the Xi cover up in place and tighten the four screws.
14.Prior to operating the O2 Probe and the Xi, all optional software enhancements (previously enabled or not) must be enabled. Notify Rosemount Analytical Inc. and reference the following part numbers to enable the related software options:
Part Number | Software Option |
6A00269G01 Enhanced Software Option Upgrade, Stoichiometric Function
6A00269G02 Enhanced Software Option Upgrade, Programmable Reference Function
6A00269G03 Enhanced Software Option Upgrade, 850° C Process Function
NOTE
For enhanced software upgrades or to enable optional software features previously used in your Xi configuration, contact Rosemount Analytical Inc. at
15.Recalibrate the O2 Probe according to the applicable calibration instructions in Section 4: Startup and Operation.