Instruction Manual

IM-106-910Xi, Original Issue November 2010

Xi Advanced Electronics

Power Supply Board Replacement

Use the procedure that follows to replace the Power Supply board in the Xi. Use this procedure to replace an original Linear Power Supply board or the current configuration Switching Power Supply board.

Disconnect and lock out power before working on any electrical components.

1.Loosen the four screws securing the Xi cover. The screws are captive and do not need to be completely removed.

2.Swing the Xi cover down to expose the inner components.

3.Refer to the wiring diagram in Figure 6-9. Unplug the AC input wiring plug from the Power Supply board. A new plug is supplied in the replacement kit and should be used if the existing plug is damaged.

4.Disconnect the 14-pin ribbon cable from the Power Supply board.

5.Remove the two long screws that secure the bracket (9, Figure 6-1) to the Xi enclosure.

6.Hold the AC input wiring to the right and slide the Power Supply board out of the Xi enclosure.

7.Install the new mounting bracket (9, Figure 6-1) on the new Power Supply board (10, Figure 6-1). A new bracket and mounting screws are provided in the replacement kit.

8.Slide the Power Supply board into the mating slots in the Xi Enclosure. Make sure the board is correctly aligned in the slots.

9.Install and tighten the bracket mounting screws. Two new screws are provided in the replacement kit and should be used if the existing screws are damaged.

Figure 6-9. Power Supply

Board Wiring

6-11

Page 73
Image 73
Emerson Process Management IM-106-910Xi instruction manual Power Supply Board Replacement, Power Supply Board Wiring