ALARM RELAY OUTPUT CONFIGURATION

Setting Test Gas Values Using Field Communicator

1.Use the Field Communicator software to access the HART menu.

2.From the DEVICE SETUP menu, select DETAILED SETUP.

3.From the DETAILED SETUP menu, select CAL SETUP.

4.From CAL SETUP, select Cal Gas 1. Enter the percent O2 used for the low O2 test gas.

5.From CAL SETUP, select Cal Gas 2. Enter the percent O2 used for the high O2 test gas.

The Xi has two dry contact Form-C alarm relay output signals that can be configured in eight different modes through the Xi keypad display or the 375/475 Field Communicator. A list of possible configurations is shown in Table 3-1. Each alarm relay output can be configured separately.

If the Xi is configured with the optional Flame Safety Interlock, Alarm 2 is configured with "Heater Relay" and prewired to the AC Relay Board. In this condition the relay configuration cannot be changed to any other setting. If the Xi is not configured with the optional Flame Safety Interlock, "Heater Relay" is not valid and cannot be chosen for Alarm 2.

Table 3-1. Alarm Relay Output Configurations

Mode

Configuration

No Alarm*

The output is not configured for any alarm condition.

 

 

Unit Alarm

The output is configured for a Unit Alarm.

Low O2 Alarm

The output is configured for a Low O2 alarm.

Low O2/Unit Alm

The output is configured for a Unit alarm and a Low O2 alarm.

Cal Recommended

The output is configured for a Calibration Recommended

 

display.

 

 

Cal Rec/Unit Alm**

The output is configured for a Unit alarm and a Calibration

 

Recommended display.

Low O2/Cal Rec

The output is configured for a Low O2 alarm and a Calibration

 

Recommended display.

 

 

Low O2/Unit/Cal Rec

The output is configured for a Low O2 alarm, a Unit alarm, and a

 

Calibration Recommended display.

Heater Relay

The output is configured for Flame Safety Interlock.

*The default configuration for Alarm 2

**The default configuration for Alarm 1

Configuring Alarm Relays with the Xi Keypad/Display

1.Press the MENU button once.

2.From the main menu, select PROBE 1.

3.From PROBE 1, select DETAILED SETUP.

4.From the DETAILED SETUP menu, select ALARM RELAY.

5.From ALARM RELAY, select as follows: Alm Relay1 - Alarm 1 mode

Alm Relay2 - Alarm 2 mode

Low O2 Alm SP - Low O2 alarm setpoint

High Temp Alm SP - High temperature alarm setpoint

6.Press the LEFT arrow key several times to return to the main menu.

7.Repeat steps 2 through 6 for PROBE 2 if configured for dual channel.