Xi Advanced Electronics

AUTOCALIBRATION SETUP

If autocalibration is desired, the Xi must be used with either an SPS 4001B or IMPS 4000. The Xi must be properly configured before autocalibration can take place. Refer to the applicable SPS 4001B or IMPS 4000 instruction manual for details on performing autocalibration. Refer to Section 4, Startup and Operation for details on manual calibration procedures.

A Xi is shipped from the factory without autocalibration configured. This same process must be performed any time a replacement I/O Board is installed.

Configuring Autocalibration with the Xi Keypad/Display

1.Press the MENU button once.

2.From the main menu, select PROBE 1.

3.From PROBE 1, select DETAILED SETUP.

4.From the DETAILED SETUP menu, select CAL SETUP.

5.From CAL SETUP, select Auto Cal.

6.Press the RIGHT arrow key to change the state from NO to YES.

Configuring Autocalibration with the Field Communicator

1.Use the 375/475 Field Communicator software to access the HART menu.

2.From the DEVICE SETUP menu, select DETAILED SETUP.

3.From the DETAILED SETUP menu, select CAL SETUP.

4.From CAL SETUP, select Auto Cal.

5.Press the RIGHT arrow key to change the state from NO to YES.

3-5