Specifications

9

Automatic Scope Measurements

Power

Power Factor

ratio between Watts and VA

Range

0.00 to 1.00

Watt

RMS reading of multiplication

corresponding samples of input A (volts)

 

and Input B (amperes)

Full Scale Reading

999 counts

VA

Vrms x Arms

Full Scale Reading

999 counts

VA Reactive

((VA)2-W2)

Full Scale Reading

999 counts

Phase

 

Range

-180 to +180 degrees

Resolution

1 degree

Accuracy

±1 degrees

0.1 Hz to 1 MHz

1 MHz to 10 MHz

±3 degrees

Temperature (TEMP)

With Optional Temperature Probe

 

Ranges (°C or °F)

............................... -40.0 to +100.0 °

 

 

-100 to +250 °

 

 

-100 to +500

°

 

 

-100 to +1000

°

 

 

-100 to + 2500

°

Probe Sensitivity

1 mV/°C and 1 mV/°F

Decibel (dB)

 

 

 

dBV

dB relative to one volt

dBm

dB relative to one mW in 50 or 600

dB on

VDC, VAC, or VAC+DC

Accuracy

same as VDC, VAC, VAC+DC