Specifications

9

Dual Input Oscilloscope

Record Length

 

Scope Record Mode

27500 points on each input

Scope Normal Mode

1000 points on each input

Scope Glitch Capture Mode... 500 points on each input

Glitch Detection

 

5 s to 120 s/div

displays glitches as fast as 50 ns

Waveform Display

A, B, A+B, A-B, A*B, A vs B

Normal, Average (2,4,8,64 x), Persistence

Time Base Accuracy

±100 ppm

Trigger Level Control Range

.........................±4 divisions

Trigger Sensitivity A and B

 

DC to 5 MHz at >5 mV/div

........................0.5 divisions

DC to 5 MHz at 5 mV/div

1 division

200 MHz (FLUKE 199)

1 division

250 MHz (FLUKE 199)

2 divisions

100 MHz (FLUKE 196)

1 division

150 MHz (FLUKE 196)

2 divisions

60 MHz (FLUKE 192)

1 division

100 MHz (FLUKE 192)

2 divisions

Trigger and Delay

Trigger Modes

Automatic, Edge,

 

External, Video, Pulse Width

Trigger Delay

up to +1000 divisions

Pre Trigger View

one full screen length

Max. Delay

10 seconds

Automatic Connect-and-View Trigger

Source

A, B, EXT

Slope

Positive, Negative

Edge Trigger

 

Screen Update

Free Run, On Trigger, Single Shot

Source

A, B, EXT

Slope

Positive, Negative

Isolated External Trigger

Bandwidth

10 kHz

Modes

Automatic, Edge

Trigger Levels (DC to 10 kHz)

................. 120 mV, 1.2 V

Video Trigger

 

 

Standards

PAL, PAL+, NTSC, SECAM

Modes

Lines, Line Select, Field 1 or Field 2

Source

A

Polarity

Positive, Negative

Sensitivity

0.7 division sync level

89

Page 97
Image 97
Fluke 196 Trigger and Delay, Automatic Connect-and-View Trigger, Edge Trigger, Isolated External Trigger, Video Trigger