Fluke 192/196/199
Users Manual

88

Dual Input Oscilloscope

Isolated Inputs A and B (Vertical)

Bandwidth, DC Coupled
FLUKE 199....................................... 200 M Hz (- 3 dB)
FLUKE 196....................................... 100 M Hz (- 3 dB)
FLUKE 192.........................................60 MHz (-3 dB)
Lower Frequency Limit, AC Coupled
with 10:1 probe......................................<2 Hz (-3 dB)
direct (1:1)............................................. <5 Hz (-3 dB)
Rise Time
FLUKE 199....................................................... 1.7 ns
FLUKE 196....................................................... 3.5 ns
FLUKE 192....................................................... 5.8 ns
Analog Bandwidth Limiters.............. 20 MHz and 10 kHz
Input Coupling................................................... AC, DC
Polarity................................................Normal, Inverted
Sensitivity Ranges
with 10:1 probe............................50 mV to 1000 V/div
direct (1:1)...................................... 5 mV to 100 V/div
Trace Positioning Range..............................±4 divisions
Input Impedance on BNC
DC Coupled..................... 1 M (±1 %)//15 pF (±2 pF)
Max. Input Voltage
with 10:1 probe......................................600 V CAT III
1000 V CAT II
direct (1:1).............................................300 V CAT III
(For detailed specifications, see Safety”)
Vertical Accuracy....................±(1.5 % + 0.04 range/div)
Digitizer Resolution...................8 bits, separate digitizer
for each input

Horizontal

Maximum Time Base Speed:
FLUKE 199.....................................................5 ns/div
FLUKE 196.....................................................5 ns/div
FLUKE 192...................................................10 ns/div
Minimum Time Base Speed (Scope Record).... 2 min/div
Real Time Sampling Rate (for both inputs simultaneously)
FLUKE199:
5 ns to 2 µs /div................................. up to 2.5 GS/s
5 µs to 120 s/div.......................................... 20 MS/s
FLUKE 196:
5 ns to 2 µs /div.................................... up to 1 GS/s
5 µs to 120 s/div.......................................... 20 MS/s
FLUKE 192
10 ns to 2 µs /div.............................. up to 500 MS/s
5 µs to 120 s/div.......................................... 20 MS/s