Displaying Recorded Data

In Normal view, the samples that roll off the screen are stored in deep memory. When the memory is full, recording continues by shifting the data in memory and deleting the first samples out of memory.

In View All mode, the complete memory contents are displayed on the screen.

4

 

 

 

 

 

Press to toggle between VIEW ALL

 

 

 

 

 

 

 

 

 

 

 

(overview of all recorded

 

 

 

 

 

 

 

 

 

 

 

 

 

samples) and NORMAL view.

 

 

 

 

 

 

 

You can analyze the recorded waveforms using the Cursors and Zoom functions. See Chapter 4: “Using Replay, Zoom and Cursors”.

Using ScopeRecord in Single Sweep Mode

Use the recorder Single Sweep function to automatically stop recording when the deep memory is full.

Continue from step 3 of the previous section:

4

 

 

 

 

 

Open the Recorder options

 

 

 

 

 

 

 

 

 

 

 

menu.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5 (2x) Jump to the Mode field.

6

 

 

 

 

 

Select Single Sweep and accept

 

 

 

 

 

 

the recorder options.

 

 

 

 

 

 

 

 

 

 

 

 

 