Fluke 192/196/199

Users Manual

Meter

Meter Input

Input Coupling

DC

Frequency Response

DC to 10 kHz (-3 dB)

Input Impedance

1 MΩ (±1 %)//10 pF (±1.5 pF)

Max. Input Voltage

1000 V CAT II

 

600 V CAT III

(For detailed specifications, see “Safety”)

Meter Functions

 

Ranging

Auto, Manual

Modes

Normal, Relative

DMM Measurements on Meter Inputs

The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C.

Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C.

General

DC Common Mode Rejection (CMRR)

>100 dB

AC Common Mode Rejection at 50, 60, or 400 Hz

>60 dB

Ohms (Ω)

 

 

Ranges

500.0 Ω, 5.000 kΩ, 50.00 kΩ,

 

500.0 kΩ, 5.000 MΩ, 30.00 MΩ

Full Scale Reading

 

 

500 Ω to 5 MΩ

5000 counts

30 MΩ

3000 counts

Accuracy

±(0.6 % +5 counts)

Measurement Current

0.5 mA to 50 nA, ±20 %

 

decreases with increasing ranges

Open Circuit Voltage

<4 V

Continuity (CONT)

 

 

Beep

<50 Ω (±30 Ω)

Measurement Current

0.5 mA, ±20 %

Detection of shorts of

≥1 ms

94

Page 102
Image 102
Fluke 199, 196, 192 user manual DMM Measurements on Meter Inputs