Fluke 2635A user manual Format, Measurement Units String, Mx+B, Vdc, Ohms

Models: 2635A

1 268
Download 268 pages 28.28 Kb
Page 140
Image 140

2635A

Users Manual

Table 4-5. Command and Query Reference (cont)

FORMAT

Response Format

 

 

 

Set the output format type to include or exclude measurement units.

 

FORMAT 1

Measurements returned without units.

 

FORMAT 2

Measurements returned with units.

 

Commands that return measurement data (like LAST?, NEXT?, MIN?,

 

MAX?) can be expressed as a number only (FORMAT 1) or as a number

 

with a measurement unit (FORMAT 2). If this command is entered during

 

scanning while logging to the memory card, an Execution Error is

 

generated. The measurement units are:

 

MEASUREMENT

UNITS STRING

 

Scaled

 

MX+B”

 

Volts DC

 

“VDC”

 

Volts AC

 

“VAC”

 

Resistance

 

“OHMS”

 

Frequency

 

“Hz”

 

Temperature °C

 

“C”

 

Temperature °F

 

“F”

With FORMAT 1 asserted, typical returns would be +890.22E+0, +230.96E-3, 072.4E+0, +003.2E; with FORMAT 2 asserted, the returns would be +890.22E+0 HZ,+230.96E-3 VAC, 072.4E+0 F, +003.2E+0 Mx+B.

FORMAT?

Query Response Format

 

Returns the output format type.

 

1

Measurements returned without units.

 

2

Measurements returned with units.

 

Commands that return measurement data (like LAST?, NEXT?, MIN?,

 

MAX?) can be expressed as a number only (FORMAT 1) or as a number

 

with a measurement unit (FORMAT 2). The measurement units are:

MEASUREMENT

UNITS STRING

Scaled

MX+B”

Volts DC

“VDC”

Volts AC

“VAC”

Resistance

“OHMS”

Frequency

“Hz”

Temperature °C

“C”

Temperature °F

“F”

Example: FORMAT? returns 2 [Measurement data will be returned with a units indicator, e.g., +230.96E-3 VAC.]

4-36

Page 140
Image 140
Fluke 2635A user manual Format, Measurement Units String, Mx+B, Vdc, Ohms