Ten Minute Tour
(continued)

xix

CANCL
ALRM
SET ALARM
1
2
Selecting Alarms. Press the ALRM key to open
the alarm menu. Each configured channel can
have two alarm limits assigned. An alarm is set
when a reading is below or above an alarm limit.
Configuration starts with an alarm limit selection,
1 or 2. Press CANCL to exit. [Figure 2-13]
DE
M
+001.00
SET Mx+B
Mx+B
CANCL
SHIFT
CANCL
RAtE
FASt
SLO
Setting Mx+B Scaling. Press the Mx+B key to
open the Mx+B menu. Up/down and left/right
arrow keys select the digits for the first parameter
(M) (default +001.00). The effect of Mx+B scaling
is to take a measurement (x) and modify it by
multiplying the measurement with M and then
adding an offset B (configured after M is set). For
example, Mx+B=+1.5x+25 applied to a
measurement of 20.000 would display
1.5(20.000) + 25 = 55.000. Press CANCL to exit.
[Figure 2-14]
Selecting the Measurement Rate. Press the
SHIFT key, release, then press the Right Arrow
key to open the RATE menu. During the
measurement portion of the scan interval, the
measurement rate can be FASt (Fast) or SLO
(Slow). The slow rate gives full 5-digit
measurement resolution, while the fast rate gives
only 4-digit resolution. The advantage of a fast
measurement rate is more readings during
continuous scanning or low scan intervals. Press
CANCL to exit. [Figure 2-12]
op79_5f.eps
Ten Minute Tour (cont)