Computer Operations 4

Computer Interface Commands and Operation

Table 4-5. Command and Query Reference (cont)

FUNC

Channel Function Definition

 

 

 

Define the Changing a channel configuration automatically erases values held in

 

review, and resets all ALARM OUTPUTS and DIGITAL I/O lines to logical high. The

 

FUNC command clears any alarm limits and scaling values for this channel;

 

therefore, define a channel function before setting alarm limits and/or scaling values

 

for that channel. measurement function and range for the indicated channel.

 

FUNC

<channel>, <function>, <range>, <terminals>

 

 

<channel> = 0, 1,2, ...20

 

 

 

 

<function> = OFF, VDC, VAC, OHMS, FREQ, TEMP

 

 

<range> = 1,2,3,...6, AUTO [VDC, VAC,OHMS, FREQ]

 

 

<range> = J, K, E, T, N, R, S, B, C [TEMP with thermocouples]

 

 

<range> = PT [TEMP with RTDs]

 

 

 

<terminals> = 2 or 4

 

 

 

Ohms and temperature measurements that use a 4-terminal configuration are limited

 

to channels 1 to 10. Select a channel function, OFF, VDC, VAC, OHMS, FREQ, or

 

TEMP. For voltage, ohms or frequency, select a range 1 to 6, as specified in the

 

table below (or AUTO for autoranging) For temperature, select a thermocouple type,

 

or PT for RTDs.

 

 

 

RANGE

VOLTAGE

OHMS

FREQUENCY

 

1

 

300 mV

300 e

900 HZ

 

2

 

3 V

3 ke

9 kHz

 

3

 

30 V

30 ke

90 kHz

 

4

 

150/300 V*

300 ke

900 kHz

 

5

 

90 mV**

3 Me

1 Mhz

 

6

 

900 mV**

10 Me

 

*300V only on channels 0, 1, and 11

** Volts DC only

The <terminals> selection is specified only when the function type is OHMS, or TEMP using an RTD. The 2-terminal selection is valid on any channel. The 4- terminal selection is valid only for channels 1 to 10 (n), which automatically clears a channel a decade higher (n+10).

Example: FUNC 9,TEMP, PT,2 [Set the function of channel 9 to temperature measurements using a Platinum RTD, and the 2-terminal connection.] Example: FUNC 5,VDC,4 [Set the function of channel 5 to volts DC, and use the 150V scale.]

4-37