2635A
Users Manual

xxii

dir
LASt
LISt
LIST
CANCL
Setting the LIST Parameters. Press the LIST
key to open the LIST menu. LIST is used to print
out all the measurements from the Review Array,
or print out a directory of all the files on the
memory card by selecting dir (Directory). To use
LIST, a printer (or PC) must be connected to the
RS-232 port. Press CANCL to exit. [Figure 5-4]
MODE
trAnS
ALAr
ALL
CANCL
Selecting the Destination Mode. ALL (All)
sends all measurement data to the destination
device (Memory Card in this example); ALAr
(Alarm) send all measurement data to the
destination device when any scanned channel is
in alarm; trAnS (Transition) sends all
measurement data to the destination device
when any scanned channel transitions into or out
of an alarm condition. Press CANCL to exit.
[Figure 5-3]
dESt
both
Print
CArd
nonE
SHIFT
FILES
ENTER
Setting the DESTINATION Parameter. Press
the SHIFT key, release, then press the FILES
key to open the MODE menu. CArd (Card)
routes data to the memory card; Print (Print)
routes data to the RS-232 connector to a printer
(or PC); both (Both) routes data to both
destinations, and nonE (None) to neither
destination. Select CArd and press ENTER.
[Figure 5-3]
HG
op79_8f.eps
Ten Minute Tour (cont)