Front Panel Operations 2

Setting Operating Conditions

A

SET Mx+B

± 0 0 0 0 0

- to -

± 9 9 9 9 9

ENTER

SET Mx+B

X . X X X X

X X . X X X

X X X . X X

X X X X . X

ENTER

SET Mx+B

x1 m

M k

Typical

ENTER

Required

(Continued from the previous page)

Selecting the B Numerical Value. Press the up/down and left/right arrow keys to enter a five digit number that defines the numerical value for the measurement offset, ignoring the decimal point or scale multiplier. The column being configured has a bright display. For example, for a B of -115, enter -11500. When the entry is correct, press the ENTER key.

Selecting the B Decimal Value. Press the left/right arrow keys to position the decimal point in the number selected in the previous step. For the example above, the setting would be -115.00. When the decimal point is correct, press the ENTER key.

Selecting the B Scale Modifier. Press the up/down arrow keys to select the desired scale modifier: X1 (x1), m (x .001), M (x1,000,000), or k (x1,000). For the example above, the -115.00 scale modifier would be x1. When the scale modifier is correct, press the ENTER key.

op26af.eps

Figure 2-14. Setting the Mx+B Scaling (cont)