Computer Operations 4

Computer Interface Commands and Operation

TOTAL?

TOTAL_DBNC

TOTAL_DBNC?

*TRG

TRIGGER

Table 4-5. Command and Query Reference (cont)

Totalizer Value Query

Returns the value of the present Totalizer count. Format of the value is +00.000E+3. When the Totalizer has overflowed, the value returned is +001.00E+9.

Example: TOTAL? returns 13.465E+3 [The present value of the totalizer count is 13,465.]

Set Totalizing Debounce

Set totalizing input debounce state, which adds a delay of 1.75 ms to each transition, allowing increased accuracy from totalizer inputs from contact closures.

TOTAL_DBNC

<dbnc_state>

 

 

<dbnc_state>

= 1 (on) 0 (off)

Use of any other value causes an Execution Error to be generated. At initial power up, totalize debounce is disabled (0). An Execution Error is generated if the totalizer debouhce setting is changed when scanning while logging to the memory card is enabled.

Totalizer Debounce Query

Returns the totalizing input’s debounce state, as selected with the TOTAL_DBNC command.

0Debounce is off

1Debounce is on SIngle-Scan Trigger

(See front of table.)

Select Trigger Type

Select the type of scan triggering option. The use of a trigger option has the same effect as pressing the front panel Qkey. An input for an external trigger is available at the ALARM OUTPUTS connector on the rear panel of the instrument, pins TR and GROUND (TRIGGER 1). Scanning can also be enabled when a monitored channel goes into alarm (TRIGGER 2).

TRIGGER 0 External trigger and Alarm trigger disabled

TRIGGER 1 External trigger enabled

TRIGGER 2 Alarm trigger enabled

TRIGGER 0 means external triggering is disabled and only normal scan interval triggering can be used. If this command is entered during scanning while logging to the memory card, an Execution Error is generated.

TRIGGER 1 means that external triggering is enabled. An acceptable low input (-0.6 to +0.8V dc) between the pins TR and GROUND on the ALARM OUTPUTS connector on the rear panel will cause the instrument to start scanning. When the TR input returns to logic high, scanning is disabled. External trigger inputs during a scan are ignored.

TRIGGER 2 means the alarm trigger is enabled. When a channel being monitored goes into alarm, the instrument starts scanning. When the channel being monitored goes out of alarm, the instrument stops scanning. The *TRG and GET commands can be used with a trigger option selection.

If the trigger type given is not one of the listed values, an Execution Error is generated.

4-55

Page 159
Image 159
Fluke 2635A user manual TOTAL? Totaldbnc TOTALDBNC? TRG Trigger

2635A specifications

The Fluke 2635A is a highly advanced multichannel data acquisition system designed for engineers and technicians requiring precision and versatility in their measurement tasks. This instrument excels in monitoring and capturing data from various sources, making it a valuable tool in laboratories, production environments, and field applications.

One of the standout features of the Fluke 2635A is its ability to support up to 20 measurement channels. Users can simultaneously monitor multiple signals, which is crucial for complex testing scenarios where numerous variables need to be tracked. The device is equipped with a 6.5-digit precision measurement capability that ensures high accuracy and resolution in readings. This level of precision is particularly beneficial in applications such as environmental monitoring, research and development, and industrial process control.

The unit incorporates a variety of measurement modes, including voltage, current, temperature, and resistance, making it exceptionally versatile. The Fluke 2635A not only handles standard signals but also integrates with various transducers, providing extensive compatibility for users engaged in diverse testing applications.

To enhance data management, the Fluke 2635A offers advanced data logging capabilities. Data can be stored internally or transferred to external devices for further analysis. The internal memory can accommodate extensive datasets, which can be crucial in long-term monitoring tasks. Integration with external software enables users to analyze and visualize data effectively, allowing for informed decision-making based on comprehensive results.

The user interface is designed for ease of use, featuring a large, bright LCD display that clearly presents measurements and status information. The device is also equipped with a simple navigation system, allowing users to switch between channels and measurement types without hassle.

In terms of construction, the Fluke 2635A is built to withstand rigorous environments. It features a robust design that ensures reliability and durability, making it suitable for both laboratory settings and field operations. Furthermore, the device supports a wide temperature range, thus enabling usage in diverse conditions.

Overall, the Fluke 2635A is a powerful and flexible instrument that provides precise measurements across multiple channels, supporting a wide range of applications. Its advanced technology, combined with user-friendly features, makes it an essential tool for professionals in various fields. The combination of accuracy, flexibility, and durability solidifies its standing as a trusted choice for data acquisition and monitoring tasks.