Computer Operations 4

Computer Interface Commands and Operation

RATE

RATE?

REMS

REVIEW_CLR

*RST

RTD_R0

Table 4-5. Command and Query Reference (cont)

Select Measurement Rate

Specifies the measurement rate. Changing the measurement rate also clears the Review array, and ALARM OUTPUTS and DIGITAL I/O lines.

RATE 0 Selects the slow measurement rate.

RATE 1 Selects the fast measurement rate.

Selection of the fast measurement rate using the RATE 1 command will speed up the measurement portion of the scan interval; however, the measurement resolution is four digits instead of five digits. For example, a reading of +115.32 with a slow measurement rate would be +115.3 with the fast measurement rate. An Execution Error is generated if the argument is not 0 or 1 or if the instrument is scanning.

Measurement Rate Query

Return measurement rate for the instrument.

0Measurement rate is slow

1Measurement rate is fast Remote without Lockout

Places the instrument in the remote mode, lights the front panel REM annunciator, and only two front panel keys are active (with special REMS functionality):

The Qkey triggers a single scan.

The Kkey returns the instrument to normal front panel control.

To return the instrument to normal front panel control with a command, use the LOCS command.

Clear Review Values

Clear all minimum, maximum, and last values (all channels) in the Review array. (It is not possible to selectively clear individual entries in the Review array.) The Review clearing operation is carried out at any time, except during the measurement portion of the scan internal. only at the completion of any scan in progress. Clearing the Review array also clears ALARM OUTPUTS and DIGITAL I/O lines to a logic high..

Reset

(See front of table.)

RTD Ice Point (R0)

For the indicated channel, store the numeric data as the RTD R0 ice point resistance, i.e., the resistance of the RTD at 32°F (0°C). Changing the ice point also clears the Review array and ALARM OUTPUTS and DIGITAL I/O lines. (The 0 portion of R0 in the RTD_R0 command is the number zero.) The default value is 100.00.

RTD_R0

<channel<, <R0>

 

<channel> = 0,1,2 ... 20

 

<R0> = 000.00 to 999.99

An Execution Error is generated if the R0 value supplied is not within the indicated range, the channel specified is invalid, the channel is defined as OFF, or measurements are active.

Example: RTD_R0 6, 124.85 [For channel 6, set the R0 ice point resistance to a value of 124.85.]

4-49

Page 153
Image 153
Fluke 2635A user manual Rate RATE? Rems Reviewclr RST RTDR0

2635A specifications

The Fluke 2635A is a highly advanced multichannel data acquisition system designed for engineers and technicians requiring precision and versatility in their measurement tasks. This instrument excels in monitoring and capturing data from various sources, making it a valuable tool in laboratories, production environments, and field applications.

One of the standout features of the Fluke 2635A is its ability to support up to 20 measurement channels. Users can simultaneously monitor multiple signals, which is crucial for complex testing scenarios where numerous variables need to be tracked. The device is equipped with a 6.5-digit precision measurement capability that ensures high accuracy and resolution in readings. This level of precision is particularly beneficial in applications such as environmental monitoring, research and development, and industrial process control.

The unit incorporates a variety of measurement modes, including voltage, current, temperature, and resistance, making it exceptionally versatile. The Fluke 2635A not only handles standard signals but also integrates with various transducers, providing extensive compatibility for users engaged in diverse testing applications.

To enhance data management, the Fluke 2635A offers advanced data logging capabilities. Data can be stored internally or transferred to external devices for further analysis. The internal memory can accommodate extensive datasets, which can be crucial in long-term monitoring tasks. Integration with external software enables users to analyze and visualize data effectively, allowing for informed decision-making based on comprehensive results.

The user interface is designed for ease of use, featuring a large, bright LCD display that clearly presents measurements and status information. The device is also equipped with a simple navigation system, allowing users to switch between channels and measurement types without hassle.

In terms of construction, the Fluke 2635A is built to withstand rigorous environments. It features a robust design that ensures reliability and durability, making it suitable for both laboratory settings and field operations. Furthermore, the device supports a wide temperature range, thus enabling usage in diverse conditions.

Overall, the Fluke 2635A is a powerful and flexible instrument that provides precise measurements across multiple channels, supporting a wide range of applications. Its advanced technology, combined with user-friendly features, makes it an essential tool for professionals in various fields. The combination of accuracy, flexibility, and durability solidifies its standing as a trusted choice for data acquisition and monitoring tasks.