2635A
Users Manual

3-10

Using Setup Load

Perform the procedure in Figure 3-6 to discard the current instrument configuration and

load a configuration saved in a previous SETUP STORE operation (Figure 3-5). A

configuration file includes channel configurations, scan interval, measurement rate,

alarms, Mx+B scaling, and temperature unit (ºC or ºF). Communication parameters,

baud, parity, CTS, and echo remain as set previously by the front panel controls. To exit

at any time (file not loaded), press the C key.

Typical
Required
FILES
ENTER
ErASE
StorE
LOAd
SEtUP
SEt99
SEtxx
SEt00
LoAd
ENTER
ENTER
Selecting the SETUP mode. Press the FILES
key to access the FILES menu. Press the
up/down arrow keys until SEtUP is displayed,
then press the ENTER key. The menu changes
to SEtUP. If an error message appears, see
Table 3-1.
Selecting the LOAD mode. Press the up/down
arrow keys until LoAd is displayed in the SETUP
menu, then press the ENTER key. The menu
changes to LoAd.
Selecting a SEtxx File to LOAD. Press the
up/down arrow keys until the SEtxx file to be
loaded is displayed in the LoAd menu, where xx
represents a number from 00 to 99, then press
the ENTER key. After ENTER is pressed, the
entire meter configuration changes to reflect the
data stored in the selected SETUP file. A display
of nonE indicates no setup files are stored.
Init
StAt
dir
dAtA
SEtUP
FILES
op42f.eps
Figure 3-6. Using SETUP LOAD to Load Configuration Files