Graco 3A0868G important safety instructions QDA1, QDB1

Models: 3A0868G

1 74
Download 74 pages 34.71 Kb
Page 57
Image 57

System Errors

Alarm and Description

Cause

Solution

 

 

 

R4

There is too little restriction in the system.

• Check that the system is fully

Ratio High Error

 

loaded with material.

The mix ratio is higher than

 

• Check that the supply pump’s

the set tolerance for an A to B

 

 

cycle rate is set properly.

component volume compari-

 

 

 

son.

 

• Check that the spray tip/nozzle is

 

 

properly sized for the flow and

 

 

application, and that it is not

 

 

worn.

 

 

• Check that the fluid regulator is

 

 

set properly.

 

 

 

 

If the alarm occurs during start up, after

Restrict gun needle travel to slow

 

purging, the flow rate was probably too

down the initial fluid delivery rate until

 

high.

fluid hoses are loaded with material.

 

 

 

 

If the alarm occurred after you were spray-

Adjust component A and B fluid sup-

 

ing for some time, the pressures from the

ply regulator pressures until they are

 

fluid supplies could be unbalanced.

about equal. If the pressures are

 

 

already about equal, verify that com-

 

 

ponent A and B dose valves are oper-

 

 

ating properly.

 

 

 

 

Slow actuation of the component A or B

Manually operate the Dose Valve A

 

valves. This can be caused by:

and B solenoid valves to check oper-

 

 

ation.

 

 

 

 

• Air pressure to the valve actuators is

• Increase air pressure. Air pres-

 

too low.

sure must be 75-120 psi

 

 

(0.52-0.84 MPa, 5.2-8.4 bar); 120

 

 

psi is recommended.

 

 

 

 

• Something is restricting the solenoid or

• There may be dirt or moisture in

 

tubing and interrupting valve actuation

the air supply. Filter appropri-

 

air.

ately.

 

 

 

 

• Dose Valve B is turned in too far.

• Refer to Valve Settings, page

 

Dose Valve A is open too far.

35, for adjustment guidelines.

 

 

 

 

• Fluid pressure is high and air pressure

• Adjust air and fluid pressure. See

 

is low.

recommended air pressure

 

 

above.

 

 

 

QDA1

Valve seal or needle/seat are leaking.

Repair the valve.

Overdose A

 

 

Slow actuation of component A or B valves.

See Ratio Low Error and Ratio High

The A dose has overshot and,

 

Error, pages 56-57.

when combined with B, is too

 

 

Running a high mix ratio and a high flow

It may be necessary to restrict the

large for the mix manifold

rate.

flow rate through the component B

capacity.

 

dose valve by adjusting its hex nut.

QDB1

 

 

 

Overdose B

 

 

The B dose has overshot,

 

 

forcing an A dose that, when

 

 

combined with B, is too large

 

 

for the mix manifold capacity.

 

 

 

 

 

3A0868G

57

Page 57
Image 57
Graco 3A0868G important safety instructions QDA1, QDB1