Calibrating and Adjusting
To adjust focus, landing, and convergence
Static Convergence
1Temporarily disconnect the power from the instrument.
2Remove the PC board shield cover from the rear of the Color CRT Module by prying evenly on all four sides.
3Reapply power.
4Display the white,
From the Color Display Test menu, select the white,
5Preset front panel brightness control, the top of the two controls to the left of the display, maximum clockwise.
6Preset front panel contrast control, the bottom of the two controls to the left of the display, to the mechanical center.
7Check the four dots which are located around the center intersection of the
V.STAT
Center dot
H.STAT
Due to interaction, the BEAM LANDING will need to be readjusted if either the H.STAT or V.STAT adjustments are made. Once the BEAM LANDING is